Ventricular Undersensing & Non Capture
The pacing spikes in this figure occur at exactly 60bpm (the baseline rate), with all spikes showing ventricular undersensing and non-capture. This should not be confused withnon-pacing, which refers to…
Easy Cardiology Learning - One Stop for ECG, ECHO, CATH & EP Cases
The pacing spikes in this figure occur at exactly 60bpm (the baseline rate), with all spikes showing ventricular undersensing and non-capture. This should not be confused withnon-pacing, which refers to…
The baseline rhythm is sinus with complete AV block. Atrial pacing (AP) is occurring, so the VDI and VDD modes can be excluded. The device was programmed in the DDI…